Description
SourceOutput
can be Normal
, Grounded
, Open
, CALOUT
, or SRC&CALOUT
.
If SourceOutput
is Normal
, the normal source output is used. This output is defined by the source mode and other source parameters.
If SourceOutput
is Grounded
, the source output connector remains grounded while the source D/A converter is internally connected to the CALOUT line in the module.
If SourceOutput
is Open
, the source remains open-circuited even when the source is started. The impedance on the output is only about 1 kilohm because the power-fail decay circuit is still connected to the output.
If SourceOutput
is CALOUT
, the source output is connected to the module's internal CALOUT line. This allows the module's CALOUT line to be driven by an external signal applied at the source output connector.
If SourceOutput
is SRC&CALOUT
, the source output is connected to the module's internal CALOUT line, and the source D/A converter is also connected to the CALOUT line. This is a combination of the Grounded
and CALOUT
values, and is useful for multimainframe calibration.
Characteristics
Vendor |
Agilent Technologies |
Usage |
AO, Channel |
Access |
Read/write |
Data Type |
String |
Read-only when running |
Yes |
Values